Surgitime Titanium SEAL (Titanium-Foil)
Surgitime
Titanium Seal (Titanium-Foil)
– CE 0434
Surgitime Titanium Seal
(Titanium-Foil) are ideal for three-dimensional bone regeneration (GBR, Guided
Bone Regeneration). If necessary, they can be secured with a matched
easy-to-handle fixation system.
Material
Titanium Grade IV (ASTM F-67) : Thickness
0,04mm.
Clinical
Procedure
The Surgitime Titanium SEAL is very flexible and
can be used for covering periodontal defects or extraction sockets. It usually
does not need fixation. The foil is trimmed to size with the edges carefully
rounded, bent to shape by prestressing and secured with Bionnovation Bone
Screws usually in the vestibulum. Foils can also be used in an exposed position
as they are impermeable so that they protect the grafting material. On
completion of treatment the pins are simply unscrewed and removed.
Comments:
The Surgitime Titanium Seal was designed for sealing
post extraction fresh sockets, and can be intentionally exposed.
Easy to handle, the
meshes are easily molded to the site to be repaired. It has a property to
acquire way, not needing to be fixed by tacks or screws. To the success of bone
formation, it is essential to promote bleeding of the socket by curettage or
decorticalization. Must be removed within 14 days. In larger defects such as
wall loss, you can remain in place for up to 30 days. In all period of repair,
the patient should undergo mouthwash or rinse with Chlorhexidine 0.12%.
Surgitime Titanium Seal has some characteristics,
such as:
- Incresed tissue
isolation;
- Auto-fastener
(not require fixing screws);
- Total occlusivity;
- Easy removal
- Used in
post-extraction defects, and isolation or occlusion of sinus buco communications;
- Extremely low biofilm retention.